Changing the dust filter bag – for maximum suction power and hygiene
The dust filter bag (item no. 5801001) should be changed if it has reached a fill level of 80%. From 80% onwards, the suction power of the drill will decrease progressively.
The drill will emit a corresponding message when the fill level reaches 60%. This gives you adequate experience to plan the change.
Bereits ab einem Füllgrad von 60% gibt das Gerät eine entsprechende Meldung aus. Sie haben dann ausreichend Zeit, um den Wechsel zu planen.
The drill emits another message when the dust filter bag is 80% full.
Replace the dust filter bag at the latest if the filter is entirely full (100%). The capacity is sufficient for up to 150 treatments.
Note
Only use original dust filter bags. Only use each dust filter bag once. Do not empty dust filter bags for reuse. Wear a suitable respirator when changing the dust filter bag. This will minimise the risk of infection from grinding dust that may contain pathogens.
Changing the dust filter bag while the drill is switched on
SYou can change the dust filter bag while the drill is switched on. However, the handpiece motor must be switched off. If the drill is switched on, the display will show messages during the process. Follow the information on the display.
If the drill is switched on, the message ‘Filter change: dust filter bag’ appears after inserting the cover. Answer the question about whether the filter has been changed.